Endoscopic intervention

The endoscopic OR is equipped with a universally applicable surgical table in the center of the room, paired with surgical lights above and a variety of imaging technologies to aid in procedures, such as colonoscopy. The room is equipped with an anesthetic ceiling supply unit at the head end, a specifically equipped surgical ceiling supply unit at the foot end.

The area of the operating room for this type of endoscopic intervention should be about 40 m². Because of the high requirement for bacteria-free air in the operating field, the operating room should be designed with an LAF outlet or individual air outlets. The room’s bacteria-free air supply is thereby ensured by terminal HEPA filters (H13/H14). Exhaust air and/or circulating air treatment should be provided for this operating room. Germ transmission into adjoining rooms is prevented by the use of negative pressure.

Isometrics

Download

PDFs and DWG files as ZIP

Log-in

Walls

Index Variop

For detailed description and index, please download project file

1. Maquet Variop wall elements

1.1. Maquet Variop illuminated glass wall element with LED technology

1.2. Sliding door, 1-leaf

1.3. Wall air exhaust

1.4. Hinged door, 1-leaf

1.5. Medical gas outlets

1.6. Clock, analog

1.7. Sliding door, 1-leaf

1.8. Control panel

1.9. Nurse workstation with 27“ monitor and keyboard

1.10. 46“ monitor

Groundplan & Ceilingplan

Index Equipment

For detailed description and index, please download project file

2.1. Mobile operating table (e.g. Maquet Alphaclassic Pro)

2.2. Anesthesia machine (e.g. Flow-i C30)

2.3. Bowl holder

2.4. Aspirator (e.g. Twista SP 1070)

2.5. Mayo instrument stand

2.6. Waste collector

2.7. Equipment table

2.8. Disposal unit

2.9. Swivel stool

2.10. Anesthesia trolley

2.11. Dual surgical light: Maquet Lucea (100/50)

2.12. Duo ceiling supply unit (e.g. Maquet Moduevo and Maquet PLG II)

2.13. OR integration system Tegris

2.14. Anesthesia ceiling supply unit (e.g. Maquet PLG II)

2.15. HEPA terminal filter units (H13/H14)

2.16. Room lights

References